Quantitative analysis of elastic recoil after balloon angioplasty and after intracoronary implantation of balloon-expandable Palmaz-Schatz stents.
1993
AbstractObjectives. The purpose of this study was to measure elastic recoil from sequential angiograms after balloon angioplasty and after implantation of a balloon-expandable Palmaz-Schatz stent in the same patient, and to compare the results with the late angiographic outcome.Background. The immediate result of coronary balloon angioplasty is influenced by plastic deformation, primarily of the atherosclerotic plaque, and by elastic recoil, primarily of the less or nondiseased vessel circumference,Methods. The extent of elastic recoil was measured quantitatively as the difference between maximal balloon size and the resulting vessel diameter or cross-sectional area.Results. Analysis was pe…
Anatomical and procedural features associated with aortic root rupture during balloon-expandable transcatheter aortic valve replacement
2013
Background— Aortic root rupture is a major concern with balloon-expandable transcatheter aortic valve replacement (TAVR). We sought to identify predictors of aortic root rupture during balloon-expandable TAVR by using multidetector computed tomography. Methods and Results— Thirty-one consecutive patients who experienced left ventricular outflow tract (LVOT)/annular/aortic contained/noncontained rupture during TAVR were collected from 16 centers. A caliper-matched sample of 31 consecutive patients without annular rupture, who underwent pre-TAVR multidetector computed tomography served as a control group. Multidetector computed tomography assessment included short- and long-axis diameters an…